PEERLESS EUROLAB 200 (Automatic Random Access Medium Throughput Clinical Chemistry Analyzer)
In Stock
PEERLESS EuroLab 200
Automatic Random Access Medium Throughput Clinical Chemistry Analyzer)
Key Features :
- Constant 200 test / hour with double reagent, typically suitable for high workload lab.
- Equipped with 4 probes and 3×4 stages on-board laundry with drying steps for each cuvette.
- Multifunctional sampling probe with external and internal mirror polish ensuring no carry over.
- Liquid level detection.
- 8-onboard wavelengths from 340nm – 740 nm with 4 optional positions covering all routine biochemistry and immunoturbidimetry assays.
- Equipped with array optical splitter , a new generation optical assembly
- 80 reagent positions / 40 reagent positions with optical barcode system
- Continuous sample loading at 40 sample positions
- 44 individually positioned high quality semi-permanent reaction cuvettes ensuring long life.
In Stock
- 100% after sales service.
- 7 Day Replacement Policy
- Hassel Free Cancellations
- Cash On delivery
- Minimum 6 Months of Shelf life (Applicable on Reagent, Chemical & Kits)
Description
EuroLAB-200
Medium Throughput Random Access Analyzer
for Clinical Chemistry and Turbidimetry
PIPETTING SYSTEM
- Multi-functional sampling probe with external and internal mirror polishing ensuring no carry-over
- Dedicated probe equipped with liquid level detection, auto probe depth adjustment and collision protection
- Separate paddle-type mixer for effective mixing
- Internal and external probe washing with jet force ensuring no carry-over
ON-BOARD WASHING
- A true walk-away system
- Equipped with 4 probes and 3 × 4 stage on-board drying steps for each cuvette
- Maintenance-free pumps for better economy
- Cuvette verification facility to ensure optical performance
REAGENT AND SAMPLE CAROUSEL
- 80 reagent positions with maintenance-free 24-hour cooled compartment for better reagent stability
- Continuous sample loading at 40 sample positions
- Automatic pre & post dilution of samples
- Compatible with primary tubes and sample cups
- Dedicated system pack reagents for superior performance
REACTION CAROUSEL
- 44 individually positioned high-quality semi-permanent reaction cuvettes ensuring long life
- Temperature maintained reaction carousel at 37°C ± 0.2°C
OPTICAL SYSTEM
- 8 on-board wavelengths from 340 nm–700 nm
- 4 optional wavelength positions covering all routine biochemistry and immunoturbidimetry tests
- Equipped with array optical splitter, a new generation optical assembly
- Integrated fully closed optical system to avoid external interference of light
UNIQUE PROPOSITIONS
Constant 200 Tests/Hour
- A true walk-away system with constant 200 Tests/hour with double reagent
- Typically suitable for high-workload laboratories
Maintenance-Free Sampling Probe
- Equipped with maintenance-free multifunctional sampling probe
- Backed by new-generation maintenance-free high precision ceramic piston-based syringe pump
Intelligent Probe Washing
- Internal & external intelligent probe washing with jet force ensuring no carry-over
Advanced Optical System
- Advanced optical fibre-based array optical splitter
- Ensures long life of optical assembly
- Helps in multi-wavelength simultaneous detection
User-Friendly Software
- Versatile and user-friendly software interface
TECHNICAL SPECIFICATION
SYSTEM FUNCTION
- System: Discrete random access automated analyzer
- Throughput: Constant 200 Tests/hour (Single / Double reagent)
- Analysis method: 1 point-end, 2 point-end, Fixed Time, Kinetic
SAMPLE & REAGENT SYSTEM
- Sample positions: 40 sample positions
- Sample & reagent probe: 1 independent multi-functional mirror polished probe
- Sample tube: Micro-cup & blood collection tubes
- Sample volume: 2–30 μL with 0.1 μL stepping
- Sample probe features: Liquid level detection, collision protection, timely tracking of liquid volume
- Reagent positions: 80 reagent positions / 40 reagent positions with optical barcode system
- On-board cooling: Non-stop on-board cooling through dual-Peltier
- Reagent volume: 20–300 μL with 1 μL stepping
- Sample & reagent probe cleaning: Automatic washing both interior and exterior
REACTION SYSTEM
- Reaction tray: 44 semi-permanent reaction cuvettes
- Cuvette specification: Individually positioned amorphous optical plastic reaction cups
- Reaction volume: 150–300 μL
- Reaction temperature: Peltier pad incubation with 37°C ± 0.1°C
- Reaction cuvette cleaning: On-board 3 × 4 stage cuvette washing with cuvette verification
MIXING SYSTEM
- Mixer: Paddle-type mixer for homogeneous mixing, carry-over reduction technology
OPTICAL SYSTEM
- Light source: Halogen lamp
- Wavelength: 8 wavelengths:
- 340 nm
- 405 nm
- 450 nm
- 505 nm
- 546 nm
- 578 nm
- 630 nm
- 700 nm
- (4 free optional positions)
- Absorbance range: 0–4 Abs
- Resolution: 0.0001 Abs
CALIBRATION AND QC
- Calibration rules:
- 9 types:
- 1-point linear
- 2-point linear
- Exponential
- Logit-log 3P
- 4P
- 5P
- Multi-point linear
- Spline
- Line graph
- 9 types:
- Control rule: Westgard multi-rule, Levey-Jennings (L-J) Plot
OPERATING SYSTEM
- Operating system: Windows XP / Windows 7 / Windows 8 / Windows 10
- Data processing:
- Automatic calibration
- Test combination
- Effective reagent management
- Blank deduction
- Dirty cuvette memory
- Cross-infection prevention
- Patient information memory and associated input
- Report statistics and printing
- Report printing: 6 optional formats, supports custom templates
- Interface: TCP/IP network interface
WORKING CONDITIONS
- Power supply: AC 110/220V ±10%, 50/60Hz, 650W
- Temperature: 15–30°C
- Humidity: 35–80%
- Water requirement: Average 4–5 L/hour under typical workload; maximum 8 L/hour at 200 tests/hour in double reagent mode
WEIGHT & SIZE
- Dimensions (L × W × H): 870 mm × 690 mm × 740 mm
- Weight: Approximately 80 kg








Reviews